Title: Extruded Tubing: Super Lubricating Additives (Complete Guide)
For medical device designers seeking lubricious surfaces for catheter devices, PTFE, FEP, and HDPE have been the materials of choice for extruded parts. As a contact layer for interface devices, these lubricious materials can be easily inserted into the human body or other devices, improving motion sensitivity and pushability. However, due to cost, performance, or process considerations, these materials are not always the best choice for specific applications. To provide medical device designers with more options, leading material manufacturers have developed low-friction additives that can be mixed with medical polymers, significantly reducing surface friction and making the extruded tubing surface more slippery. These additives can be used with most existing medical polymers without significantly affecting the original properties of the main material. These lubricious materials can be used:
– As a lining for tubing
– As an alternative to hydrophilic coatings
– To improve adhesion to other components
**Purpose of the Article:**
The dilemma for medical device designers is which lubricious material or additive is suitable for their project. Due to a lack of market data, this can be a challenge. Data provided by lubricant additive manufacturers is often limited to how their product affects the lubricity and mechanical properties of the medical tubing it’s mixed with. However, test results from different manufacturers cannot be accurately compared.
The technical staff at Micro-tube conducted independent tests to understand the performance of several leading lubricant additives in common extruded medical polymers. Tables 1 and 2 list the selected additives and materials, respectively. We tested these extruded materials for:
– Extrusion capability
– Dimensional stability
– Lubricity
– Elongation
– Tensile strength
– Adhesion strength
We also tested the lubricant additives to determine the effects of aging, sterilization, and printing processes. The article summarizes the test results.
**I. Materials**
01. Selected Additives: As shown in Table 1, we tested four market-leading additives produced by four manufacturers.
02. Material Selection: All lubricant additives in Table 1 can be combined with a variety of common engineering thermoplastics, including nylon, Pebax®, TPU, polyethylene, polypropylene, PET, ABS, and polycarbonate. We compared these lubricant additives with two lubricious fluoropolymers produced by Daikin:
– NEOFLON™ EFEP RP-5000
– NEOFLON™ PFA AP-210
These fluoropolymers share some characteristics with polytetrafluoroethylene (PTFE) but have several advantages: they are easier to extrude, easier to bond, and can be gamma-irradiated for sterilization.
**II. Methods**
The staff at Micro-tube used a traditional ¾” extruder to extrude the materials in Table 2 (polytetrafluoroethylene was extruded from an external source). First, we extruded the natural materials and recorded all extrusion settings and parameters as a baseline. Second, we extruded the materials with additives using parameters as close as possible to the natural materials to minimize process changes that could affect the test results. Where possible, we used the processing conditions recommended by the manufacturer in the settings. Our results are based on a limited number of trials and are intended to provide preliminary guidance for material selection (for specific applications, we will work with customers to analyze and optimize the process to meet their needs).
**III. Results**
We observed that there was almost no difference in the extrusion process between the materials with additives and the natural materials. Tecoflex showed slight instability during extrusion, but this was expected due to the material’s natural tackiness.
**IV. Dimensional Stability**
We conducted dimensional stability tests on each material in Table 2, with a sample size of 20 and an AQL of 2.5%. We used an offline laser scanning micrometer (Scantron XLS 35XY) to measure the outer diameter (OD) size capability. The results are given in the form of outer diameter tolerance, which is the range of 20 test samples divided by 2 or (maximum value – minimum value) / 2.
**V. Mechanical Testing**
01. Lubricity: A professional external testing laboratory conducted lubricity tests for Micro-tube. The sample size was 20, with an AQL of 2.5%. The test fixture was mounted on the base of the tensile testing machine. This was a wet test conducted at 37°C using deionized water. The samples were clamped in the test fixture. A tensile testing machine pulled the samples horizontally through the fixture at a speed of 3 inches per minute. The force required to pull the samples 75 millimeters was plotted against the pulling distance. The tests were conducted using silicone (GP 60W) fixtures and PTFE fixtures.
**VI. Conclusion**
The study results show that compared with natural materials, lubricant additives and materials can be extruded on traditional extrusion equipment without obvious stability or processing issues. The data show that the dimensional stability of lubricant additives and materials is comparable to natural materials, and even better in some cases. However, the mechanical properties of materials with additives may be affected in some cases. We recommend that designers consider the key mechanical requirements of the end device before selecting materials to avoid affecting important design features. The test results also show that compared with natural materials, all additives can reduce friction. This is more evident in some additives. When tested with PTFE fixtures, the performance of PFA materials is almost comparable to PTFE materials, and EFEP materials perform even better. These materials can prove to be a good choice for designers to increase lubricity in applications where PTFE cannot be used. We also found that all materials can be bonded, printed, aged, and gamma-irradiated. In these tests, some additives and materials performed better than others. Therefore, we again recommend that designers carefully consider the process that the final device will adopt when selecting materials. From the results of this study, we believe that these lubricant additives and materials provide medical device designers with more options and flexibility in developing new and improved medical devices.
**VII. Appendix**
01. Dimensional and Mechanical Test Results Table: As described in Section 3.3.3, Table 5 lists all the test results of dimensional stability, elongation, lubricity, tensile strength, and adhesion strength for all materials.
**VIII. About EverGlide MED**
EverGlide MED series polymer compounds provide **best-in-class friction control and lubricity** while maintaining physical properties and medical compliance. EverGlide MED composites use our proprietary additive technology to effectively reduce the coefficient of friction while maintaining key performance and processability. The resulting compounds are easy to process on traditional equipment and provide a permanent, non-migrating solution for high lubricity. EverGlide MED has a wide range of applications and can be used with almost any polymer, and can be combined with pigments or fillers such as barium sulfate, tungsten, and bismuth oxide. Polymers such as polyolefins, TPU, PEBA, nylon, fluoropolymers, PPSU, PEI, and PEEK can be modified with EverGlide MED technology. Medical devices produced with EverGlide MED can be ETO and steam sterilized, and the products comply with ISO 10993 and USP Class VI standards. EverGlide MED compounds are specifically designed for medical tubing, cavities, and catheters to improve lubricity and reduce draw force, thereby eliminating the need for hydrophilic coatings. Testing was conducted in a third-party laboratory. This is a wet test conducted at 37°C using deionized water. A tensile testing machine pulled the samples horizontally over the fixture at a speed of 3 inches per minute. The data is the average force required to stretch the samples by 75 millimeters.
**Micro-tube provides medical plastic particle modification, please contact us immediately if needed!**
– **Lubrication modification**
– **Contrast modification**
* * *